Treatment principle:
Spine Stimulator, combined with medium frequency electrical stimulation and
thermal therapy, provides a dynamic, wave-like stimulation, eliminating the
time-consuming electrode positioning in the multi-channel stimulator, and the
dynamically changing waveform can make the patient experience tapping, massage
and kneading, makes the stimulation deeper and opens up a new way of treatment.
It incorporates the characteristics of medium frequency and low frequency. The
modulated medium frequency current contains medium frequency electrical
components, so it has the characteristics of medium frequency. The human body has
low impedance to it and has a deeper effect, stronger current can be used; no
electrolysis, no skin irritation , can give full play to the unique physiological and
therapeutic effects of medium frequency current. Modulated medium frequency
current contains low frequency components, so it has the characteristics of low
frequency current at the same time, which can exert the physiological and
therapeutic effects of low frequency current.
Technical parameters:
1. The electrode pads is designed according to the shape of back of human, which
can fully contact the back of the human body for better treatment effect.
2. The treatment site can be precisely controlled.
3. 8-inch LCD touch screen display and control.
*4. Built-in prescriptions: 77, including 19 neck-shoulder mode, 19 chest-back mode,
21 waist mode, and 18 combination mode.
5. Carrier Waveform: under the condition of 500Ω load impedance, a two-way
symmetrical rectangular wave.
6. Carrier frequency: 2000Hz, error±10%.
7. Pulse width: 250μs, error±10%.
8. Output amplitude: the output amplitude is adjustable from 0 to 35V, divided into 0
to 99 levels, error±10%.
*9. 15 kinds of waveforms.
10. Output current: the maximum output current is 60mA, error±20%.
11. Temperature: heating temperature setting range: 37℃~41℃ adjustable in
five levels, error±3℃.
12. Continuous working time: ≥4h.
13. Case storage: store 40 patient information, the storage content is name,
gender, age and remarks.
14. Output current stability: The output current change rate should not be
greater than 10%.
15. Timing range: adjustable 1min-30min, with a step difference of 1min, default
value 20min, error±30s. After the treatment time is over, the output will stop
and an audible prompt will sound.